The microfluidic chip Fluidic 719 consists of a flow focusing droplet generator connected to a channel with 2261 microfluidic droplet traps of 173 μm diameter. The microfluidic traps work by floatation, profiting from the difference in density of the two phases of an emulsion. If droplets of the same size as the traps are produced, they will remain individually trapped over time and can be identified according to their position. Combined with the multi-injection valve MUX distributor, the dispersed phase can be seamlessly changed to produce batches of different droplets in the same chip to be observed and compared under the same conditions. This Application Note describes the procedure and equipment involved to achieve sequential production of different droplets and their storage in microfluidic traps for in situ optical analysis.
Flow controller OB1
Flow sensor
MUX Distribution Valve
Fluidic 719 – ChipShop
Figure 1: Setup schematic
There is one flow focusing device per chip to produce the droplets to be trapped. There is one inlet for the continuous phase (1) and up to three inlets (2-4) for the dispersed phase (Detail A). Two of these can be used to mix different substances right before droplet formation, however, in this application we will only use one of them to generate a variety of different droplets. There are two other inlets in the opposite side that can be used to produce double emulsions that will not be used in this experiment (6&7). Once produced, the droplets are transported though a long serpentine channel with seventeen straight sections. Each one of these sections has 133 droplet traps of 173 μm diameter. The traps are positioned alternatively in a zigzag pattern along the channel, allowing flow around the trap if occupied by a droplet. Therefore, the droplet size is critical in this experiment, as bigger droplets will not be trapped and smaller droplets will be pushed away. By making droplets in the 150-180 μm diameter range, we ensure there will be only one droplet per trap and it will remain stable in that position. There are two different outlets, one right before the serpentine channel that can be used to extract droplets after production (8) and another one at the end of the serpentine channel (5). Connections to the chip is done using Microfluidic ChipShop Mini Luer connectors, while the inlets and outlets that are not in use are sealed using low volume displacement male Luer plugs.
Tips from the expert. These parameters are adapted to the oil used and dispersed phase with the properties of water using 20 cm of 100 µm ID resistance tubing. If phases of different properties, viscosities or different resistance tubing are used, these values should be adjusted for reliable flow control.
Tips from the expert. For more details on the use of resistance to obtain the best performances in terms of flow rate control please refer to the “Flow control tuning” document.
6. Secure another piece of 1/32” tubing to a waste reservoir (using a cap, a piece of tape…) and connect the opposite end to a mini Luer male fitting.
Tips from the expert. The chip is connected in later steps to make sure it is completely filled with oil before the dispersed phase sequence reaches the flow focusing nozzle. This will aid generation of stable droplets of the right size and sequence, as well as limiting air bubbles.
The following steps detail the formation of a sequence of aqueous solutions in the tubing, separated by small plugs of oil to prevent them from mixing, before the droplet generation chip. Preparation of the dispersed phase sequence before connecting the tubing to the droplet generation and storage chip aids in the downstream sequential production of droplets with different contents because of the very small volumes involved.
Tips from the expert. The channels and resistance should be filled with liquid to obtain an accurate reading of the flow rate and use flow control as opposed to pressure control. Therefore, the liquid needs to have the properties of the dispersed phase corresponding to water, as specified in the sensor parameters.
5. Set pressure (200 mbar for example) of the dispersed phase (MFS2) until the solution starts dripping out of the end of the tube connected to the mini Luer male fitting (i.e. that will be connected to the chip inlet later).
6. Switch the control settings from pressure control to flow control (“sensor” toggle) and set the flow rate to 0.5 μl/min. Then, prepare the sequence in the MUX distributor interface.
7. Select the reservoir containing the oil from the MUX distributor window, and let it flow at 0.5 µl/min for 1 minute to generate a 0.5 μl oil plug. The oil will act as a spacer to prevent the subsequent solutions from mixing inside the tubing.
Tips from the expert. The volume of the oil plug is not crucial and does not need to be exhaustive. However, bigger plugs will considerably affect the flow control and smaller plugs can fail to reliably separate consecutive plugs of reagents inside the 1/32” OD tubing.
8. Sequentially change the MUX distributor valve position according to the sequence of droplets to be produced later. Each of these aqueous solutions must be separated by 0.5 μl of oil to avoid cross-contamination of the dispersed phases.
Tips from the expert. The ideal droplet size is equivalent to 1.5-2 nl of volume. This means that the total volume of dispersed suspension for 2261 droplets, not counting oil spacer volume, is approximately 4 μl. Thus, to prepare a dispersed phase sequence of 4 different aqueous reagents, ~1 μl of each reagent is required. Using a constant dispersed phase flow rate of 0.5 μl/min, prepare an injection sequence where each reagent flows for 2 min. See the Sequential Injection User Guide for tips on creating injection sequences.
9. The whole sequence should be finished by another plug of oil. Many sequences can be prepared to fill multiple chips consecutively. Depending on the stability of reagents, even pieces of tubing containing the desired sequence can be produced, sealed and stored to later produce the droplets as long as the air is expelled before connecting the chip
Tips from the expert. In case a previously prepared sequence needs to be injected, just fill the whole circuit after the MUX distributor with neutral solution and then substitute the tubing piece at the end, connected after the microfluidic resistance with the prepared piece.
10. Select an aqueous solution with the MUX distributor and use it to push the sequence through the tubing connected to the MUX outlet until the first plug of oil reaches the end of the tube connected to the mini Luer male fitting.
Tips from the expert.At this stage, the flow rate can be changed to up to 4 μl/min to accelerate the waiting time. However, higher flow rates might affect the prepared sequence and the flow should be progressively slowed down as it nears the end of the tubing.
11. Stop the flow, and make sure the sequence stays still in the tube by placing it gently on a flat surface.
12. Set pressure (200 mbar for example) of the continuous phase (MFS3) until the solution starts dripping out of the tubing and then connect the mini Luer male fitting to inlet 1 of the Fluidic 719 chip.
13. Connect the mini Luer male fitting from the waste to outlet 5.
14. Seal all remaining inlets and outlets with low volume displacement plugs.
Tips from the expert.Other microfluidic plugs can also be used. However, once all traps are filled, big volume fluctuations may push or displace the trapped droplets.
15. Fill the microfluidic chip with oil and switch the control settings from pressure control to flow control to a set flow rate of 25 µl/min.
16. Remove the plug from inlet 3 and connect the mini Luer male fitting from the MFS2.
17. Restart the flow of the dispersed phase, now connected to the chip, at 0.7 µl/min.
18. The droplets will then be produced sequentially. It is convenient to supervise the droplet production process and adapt the dispersed phase flow rate in case of fluctuations in droplet size.
Tips from the expert. The oil flow rate is carefully chosen in tune with the dispersed phase for droplet production and to optimize droplet trapping. A faster flow rate might push droplets out of the traps, while a slower flow rate might cause droplets to get stuck outside the traps, where they are squeezed and not stable over time. If any of these happen, the flow rate can be adjusted, but it is recommended to stay within the 20-30 µl /min range. If the droplets move too slowly even at such flow rates check for leaks in the circuit.
19. Once the sequence is finished, stop the channel corresponding to the dispersed phase (MFS2), and after a minute, substitute the corresponding mini Luer male fitting for a low displacement mini Luer plug.
20. Keep the flow rate of the continuous phase stable and running until all droplets are stabilized in a trapped and none of them are circulating along the serpentine channel.
21. Remove the mini Luer male fitting from the outlet and leave the outlet unstoppered (open).
22. Decrease the flow of the continuous phase progressively until it reaches zero.
23. Carefully remove the corresponding last mini Luer male fitting (from port 1) and quickly seal both the inlet (port 1) and the previously unplugged outlet (port 5) with low displacement mini Luer plugs.
Tips from the expert. The oil used in this application note evaporates fast, hence the need for sealing the inlets and outlets of the chip as soon as possible when they are not in use. If air or a bubble reaches the serpentine channel, it will likely displace the droplets due to its lower density and ruin the sequence.
Tips from the expert.In the case of cell culture, given the affinity of oil for oxygen, the droplet environment will remain aerobic for a few hours. Dynamic culture can be done by continuously circulating oil, but this will progressively reduce the droplet size over time.
24. The chip is now ready for observation and will be stable for days.
Droplets were generated using the droplet generation and storage chip (Figure 1). A quick indicator of the right size of the droplets (150-180 µm diameter) can be visualized right after production. Immediately after the dispersed phase is pinched off, the droplet should be circular and almost tangential to the channel walls.
Figure 1. Sequential droplet generation viewed by bight field microscopy. Droplets are generated at the flow focusing junction of the chip. The dispersed (aqueous) phase is entering from the left. The continuous phase (oil) is entering perpendicular to the dispersed phase from both sides (visible left hand side of image) to steadily pinch the dispersed phase into droplets
Droplet contents were changed by preparing sequentially alternating aqueous solutions using a rotary valve positioned before the chip. Following this protocol, the liquids do not mix inside the tube or the in the nozzle during droplet production. Oil plugs between the aqueous solutions become part of the continuous phase when they meet at the nozzle of the droplet generator. Thus, there is no transition gradient between aqueous reagents and droplet boundaries are sharp (Figure 2).
In the case that reagents cannot be differentiated under the microscope (e.g. if they are the same color), a reservoir of colored water can be connected to the MUX distributor and used to indicate the end of one reagent and start of the next. In order to create a spacer of colored droplets, instead of using only a plug of 0.5 μl of oil for separation, a sequence of 0.5 μl oil, 0.5 μl colored water, 0.5 μl oil should be used.
Figure 2. Stabilized droplets trapped inside the microfluidic chip. The limit between four different solutions (here shown as droplets made from green, red, blue and yellow dye) was sharp and easily discerned.
Droplets are stable in the chip for days for in situ analysis.
Application note written by Jesús Manuel Antúnez Domínguez
This application note is part of a project that has received funding from the European Union’s Horizon 2020 research and innovation programme under grant agreement No. 812780.
How can we help you?
Name*
Email*
Message
Newsletter subscription
We will answer within 24 hours
By filling in your info you accept that we use your data.
Do you want tips on how to best set up your microfluidic experiment? Do you need inspiration or a different angle to take on your specific problem? Well, we probably have an application note just for you, feel free to check them out!
A protocol for high monodispersity and high encapsulation efficiency double emulsion production for encapsulation using microfluidics
Droplet-based microfluidics and single spore encapsulation offer the key for a breakthrough in antifungal screening and fungicide discovery.
Generate picoliter-sized microdroplets and perform single cell encapsulation with this detailled microfluidic protocol.
Nanobubble generation performed by microfluidics is described in an application note. Nanobubbles are very singular by their generation and properties.
Everything you need to know about microbubble generation, from theory to the microfluidics experimental steps.
This application note will show you how highly monodispersed droplets can be easily generated to encapsulate single spore of fungus using a microfluidic droplet generation system.
Highly monodispersed alginate beads can be easily generated with a microfluidic droplet generation system.
Two-phase flow microfluidics allow to perfom different laboratory functions on one microfluidic Lab-on-Chip (LoC) platform. An example of such an application is the controlled production of droplets on chip.
The generation and manipulation of droplets through microfluidics offer tremendous advantages: better control over small volumes of fluid, enhanced mixing, high throughput experiments.
This application note describes how to generate controlled size millifluidic droplets in a capillary by regulating the pressure rate and/or by regulating the flow rate.
Droplets generation has a large scale of applications, such as emulsion production, single cell analysis, drug delivery or nanoparticles synthesis. Droplets can also be used as micro bioreactors for chemical or biochemical reactions.
Precise and effective control of droplet generation is critical for applications of droplet microfluidics ranging from materials synthesis to lab-on-a-chip systems.
Active droplet generation in microfluidics is of high interest for a wide range of applications. It provides an additional degree of freedom in manipulating both the size and the formation frequency of micro-droplets. This additional control is extremely desirable for complex operations which rely on the accurate control of both parameters.
Microfluidic droplet in a chip also called droplet generation by microfluidics presents many application & advantages. The droplet generation pack has been designed to fit most common droplet generation needs of researchers.
Droplet microfluidics and single cell encapsulation offer the key for a discovery engine in antibody therapy.
Get a quote
Collaborations
Need customer support?
Serial Number of your product
Support Type AdviceHardware SupportSoftware Support
Subject*
I hereby agree that Elveflow uses my personal data Newsletter subscription
Thanks for your interest
Get this Application Note in PDF, right in to your mailbox!
I want to receive the Elveflow newsletter
Message I hereby agree that Elveflow uses my personal data Newsletter subscription