Microfluidic technology is creating powerful tools for cell biologists to control the complete cellular microenvironment, leading to new questions and new discoveries. We review here the basic cell biology concepts and methodologies in designing microfluidic devices, and their diverse cell applications.
.
Microfluidics is a well-understood domain of physics and can now be used to develop tools for cell biology. By simply miniaturizing macroscopic systems and taking advantage of the possibility of massive parallel processing, some microfluidic chips enable high-throughput biological experiments. Specific effects of laminar flow at the micron-scale also allows spatial control of liquid composition at subcellular resolution, fast media changes and temperature changes, and single cell handling and analysis. Microfluidic technology enables studies of cell behavior from single- to multicellular organism levels with precise application of experimental conditions unreachable using macroscopic tools.
In the 80’s and 90’s, microfluidic devices were mainly fabricated on silicon substrates. These technologies required clean room facilities and strong know-how. In the late 90’s, the introduction of soft lithography using molding of polymers allowed the fabrication of cheap microfluidic devices with additional advantages due to the physical characteristics of those polymers.
The current most popular technology for the fabrication of microfluidic devices for cell biology application is based on the soft lithography of poly-di-methyl siloxane (PDMS). PDMS is an elastomer which, through simple molding procedures, can be turned into microfluidic devices. Its wide use as a material of choice is due to its mechanical properties, which make it amenable to the integration of fluidic valves, essential elements for major microfluidic applications. Further, PDMS is transparent, biocompatible, and permeable to gas, which explains the strong interest of the scientific community in using this material to fabricate microfluidic devices for cell biological studies.
The development of soft lithography gave a simple technology to fabricate devices that integrate channels at the scale of a cell. In most cases, the interest of biologists for microfluidics did not stem from an interest in new physical phenomena at the microscale, but instead from a practical experimental point of view and a favorable scaling of physical forces. At the microscale the laws of physics remain the same as in macroscopic systems, but the scale factor can give predominance to different forces. For example, in the case of fluid flow, the reduction in size reduces the influence of the inertial forces compared to frictional forces, leading to the formation of laminar flow in microfluidic channels. Furthermore, the reduction of size has a direct influence on the characteristic times of the system, such as the time required for the diffusion of a molecule, which decreases as the square of the characteristic length. Microfluidic devices give several advantages for cell biology applications. Some advantages come from the fast response of microsystems. The fast diffusive heat and mass transfer at the microscale (microscale characteristic times are approximately 10-3s-1s, compared to macro-scale time of 102s-104s) allows for fast media and environmental changes and fast temperature control. Laminar flow properties are also useful since they enable the formation of static and dynamic gradients at subcellular resolution.
Microfluidics also have a number of other positive practical aspects like low reagent consumption (nL), the opportunity to manipulate large number of cells simultaneously and independently, automatic generation of a large number of different individual conditions, and easy integration of numerous analytical standard operation and large-scale integration.
From cell biology technological point of view, soft lithography enables the integration of subcellular scale physical and chemical patterns to study cell behavior under a large spectrum of parameters. In addition, electrode integration inside the microfluidic devices can generate large localized electric fields using small voltages. Finally, the versatility of these devices partly enables the simulation of in vivo cellular microenvironments (vascularization, 3D, nutrient stress, etc…).
Microfluidics have some drawbacks. For example, laminar flows only produce relatively slow diffusive mixing, which can be a major limitation for some applications requiring fast homogenization of flow. This can be corrected using different types of integrated mixers [1], such as advective mixing in a microchannel [2]. In addition, the small reagent consumption theoretically reachable in microfluidic devices is generally not reached due to a current lack of methods for fluid handling. Changes in scaling can give further difficulties in the adaptation of biological protocols to fit experiments in microsystems (i.e., media and cell concentrations). Furthermore, PDMS has an affinity for small hydrophobic molecules and thus could lead to biomolecule absorption/adsorption from the medium, thus biasing the experimental conditions. The permeability of PDMS to water vapor can also lead to media drying and thus change its osmolarity. These differences require careful comparison between the data obtained in macroscopic experiments and data obtained in microsystems.
In this review we will discuss the use of microfluidics to fabricate research tools in cell biology with a particular focus on PDMS soft lithography. We will first describe technologies used to fabricate PDMS microfluidic devices. Then, we will describe the existing microfluidic tools in different categories: environmental composition control, mechanical deformation, force measurement, temperature control, and integration of electric fields. Finally, we will describe the main applications of these tools in cell biology and discuss the choices of PDMS for these applications.
We describe here the two major techniques, often referred to as soft lithography of PDMS, for fabricating microfluidic devices dedicated to cell biological research and for printing molecular micropatterns on cell culture substrates: PDMS casting and microcontact printing.
Figure 1 illustrates a typical procedure for making a microfluidic device. PDMS microfluidic devices are generally fabricated using molding methods [3, 4]. A silicon master mold, containing a photoresist pattern representing the channel design, is fabricated using photolithography. Once the master mold is fabricated, if carefully used and properly treated with an adhesive, it can be used hundreds of times to mold a PDMS replica of the channels.
Photolithography processes for mold fabrication, which will not be detailed here, require a spincoater and a dedicated UV lamp. For most of the biological applications described in this review, including patterned features as small as 1-micron, the photolithography equipment can be installed in a classical biological fume hood instead of a clean room, thus significantly reducing costs.
To fabricate a monolayer PDMS microfluidic device, liquid PDMS and a curing agent are mixed and poured onto the master mold and cured at 60°C for 2 hours. Then the mold and PDMS replica are disassembled, cut to the desired dimension before drilling of injection holes using generally a syringe needle. The PDMS microfluidic device and glass cover slip are plasma treated for 30s, and put into contact to produce a covalently bonded and sealed full microfluidic chip.
Since one PDMS layer can be bonded onto another using a similar plasma treatment, this technology enables the fabrication of multilayer microfluidic devices. The possibility to fabricate multilayer devices, coupled with the low Young’s modulus of PDMS, enables the integration of microwaves [5]. The integration of elements like valves gives active control of the liquid inside the chip, leading to fully automated microsystems.
Applications requiring the integration of electrodes or resistors require the deposition of metal and dielectric layers inside the device. Integration of electrodes on a substrate is a common operation in microtechnology but requires clean room facilities. Furthermore, due to the low surface energy and high softness of PDMS, metal deposition is difficult to achieve on PDMS surfaces. Nevertheless, since glass substrate is compatible with most of the clean room techniques developed for silicon-based microelectronic industry, it is possible to coat metallic and dielectric layers directly on the glass substrate prior to plasma-bonding onto the PDMS device.
Figure1. Fabrication procedure for a single layer microfluidic device.
(See text for details.)
The microenvironment of the cell is defined by chemical and mechanical parameters. Chemical environment is composed of soluble molecules around the cell which are related to the cell culture medium composition, and mechanical environment is composed of the extracellular matrix (ECM) which is related to the culture substrate composition. In conventional cell cultures, these environmental parameters are easily controlled for a population of cells, but cannot be addressed locally to individual cells. One of the major interests of microfluidics is environmental control at the scale of the cell. In addition to the ability to locally address parameters of the cell microenvironment, microfluidics also offers the ability to change these parameters dynamically and automatically due to the speed of the physical processes at the microscale and the different automation possibilities. Although an ideal comprehensive microenvironmental control device does not exist, individual environmental parameters can readily be controlled using soft¬lithography and in some cases can be combined. We will review in the next part the different existing methods to control the chemical and the mechanical microenvironment of the cell.
One of the first applications of microfluidics in cell biology was to control the cell medium. There is a strong interest in producing chemical gradients to mimic natural stimuli which occur in biological processes such as cell migration, differentiation, or development. The study of cellular response to chemical gradients requires fine spatial control of local concentrations because cells can respond to concentration gradients localized in a region as small as 2% of their diameter [6]. The poor spatiotemporal resolution of macroscopic gradient generators (MGGs) led to an interest in fabricating microscale gradient generators (μGGs). Microfluidic devices can create multiple biochemical gradients with controlled spatiotemporal distribution and subcellular resolution. These microfluidic devices offer fast response times to study fast responsive systems such as immune cells. μGGs have been successfully used to study neural stem cells growth and differentiation [7], neutrophil chemotaxis [8-10] or migration [11], bacteria chemotaxis [12], endothelial cells migration [13], cancer cells chemotaxis [14], cellular response to viruses [15], and yeast gene expression under gradients of pheromones [16]. For more information, we refer the reader to the excellent review of Keenan and Folch [17]. Study chemotaxis using the chemotaxis Pilot Pack.
MGGs have traditionally been developed using hydrogel made from fibrin, collagen or agars. Gradients are deposited on the hydrogel using an array of droplets containing biomolecules which then diffuse in the gel to form a gradient [18, 19]. A second method uses micropipettes to inject biomolecules into the gel at controlled rate to generate a gradient [19]. In general, hydrogel enables easy gradient production, but has poor reproducibility and spatiotemporal control over the gradient. In addition, the opaque optical properties of the gel can be a limitation for some applications. Other MGGs have been developed using chambers separated by membranes [20-22]. These methods cannot generate complex and stable gradients. An exception is the Dunn chamber [22] using modified micropipette techniques, which were able to generate stable gradients for several hours. Although these MGGs have helped address numerous questions, they are not useful for studies requiring gradients with precise spatiotemporal control and reproducibility.
Microfluidics enables the creation of a large spectrum of gradients: time invariant gradients, subcellular resolution gradients, continuous or discrete gradients, fast response dynamic gradients. Most of the μGGs described here use simple technologies (i.e., single layer PDMS), and have been applied to generate concentration gradient of diffusible molecules to study bacterial chemotaxis [12], and cell migration in response to chemokines [13], and to generate surface gradients of adsorbed ECM molecules to study the dependence of axon growth of neurons on the surrounding ECM composition [23].
Figure 2 gives examples of gradient generation. One simple way to generate gradients in microfluidic systems is to use the properties of laminar flows. Laminar flow-based µGGs use diffusive mixing between two or more parallel laminar streams of different composition to generate molecular gradients. The shape of the gradient based on laminar flows depends on the flow rate and the time the streams are in contact. Gradients generated in these types of devices will maintain their shape at constant flow rate. The simplest μGGs of this type is the T¬sensor [24], which is composed of two microchannels. These gradient generators have a small time constant and are theoretically able to establish or modify a gradient in 10-2s-10s. T-sensor devices are easy to fabricate and to describe mathematically. The constant perfusion permanently exchanges the medium and prevents the accumulation of cell waste products, thus enabling long cell culturing time. In contrast, T-sensor devices are reagent-consuming. They remove the autocrine/paracrine or other secreted signals of cultured cells, and submit cells to shear stress due to flow. Further, the useful region is limited to a short portion of the channel which generates only sigmoidal-shaped gradients in the direction perpendicular to the flow. To overcome this limitation, additional inlets can be added to generate more complex flow profiles.
Yet, the addition of needed controls can rapidly make these experiments tedious. T-sensor devices have been used in studies of bacterial chemotaxis [12], or endothelial cell migration [13].
An upgraded version of the T-sensor called premixer µGG [25] splits and recombines inlet fluids before merging them in the culture channel, and thus can generate more complex gradients (e.g. sawtooth and hill). By adjusting the inlet flows, premixer µGGs can generate smooth or step gradients. Dertinger et al. introduced the multiple premixer arrays which can generate overlapping gradients [26]. This type of device has been used in studies of neutrophil chemotaxis [8, 9], neural stem cell differentiation [7], and breast cancer cell chemotaxis [14]. Except for the wider range of gradient shapes, the premixer device shares the same issues as the T-sensor.
Another upgraded version of the T-sensor called universal µGG [27] includes a series of walls to split the streams. This configuration can generate many profiles of concentrations, and can reduce dead volumes compared to the T-shaped µGG, but also shares the same issues as the T-sensor and is more challenging to describe mathematically. Cookset et al. developed a μGG composed of an array of 16 multiplexed inlets, which gives 64 combinations of chamber feeding [28]. This device integrates a mixer which can be turned on and off with a bypass valve, and can produce gradients or homogenized mixtures. This device enables simultaneous formation of complex gradients of different biomolecules, with sub-second temporal resolution.
Major drawbacks of all laminar flow μGGs are that they require precise control of the flow rate. The shear stress produced by the flow can change the migratory behavior of the cells [29], and produce undesired mechanical stress on the cells, and flush away important factors secreted by cells [17]. Although possible, these µGGs are challenging for studies of nonadherent cells such as yeast or bacteria because of the movements generated by the flow [30].
The second type of μGGs is not based on the properties of laminar flow. The flow resistive µGG uses flow resistive elements to eliminate convection around the cells. This kind of device allows passive diffusion of biomolecules through a flow barrier to generate gradients. The flow barrier can be a hydrogel [31-33], nanopore membrane [10], or microchannel [32]. Hydrogel completely eliminates convection, whereas microchannels (which are easier to integrate) only minimize convection. In both cases, these devices can generate steady-state gradients, eliminate shear stress generated by flow, and preserve the autocrine/paracrine signals secreted by cells. In addition, they use less reagents than laminar flow μGGs, are possible for experiments with nonadherent cells like yeast [16], and some are able to generate gradients in hydrogel for 3D cell culture. Major drawbacks of these μGGs are their inability to create complex profiles, their large time constant compared to laminar flow µGGs, and, for the hydrogel-based µGGs, the greater difficulty to fabricate them than single PDMS layer-based laminar µGGs.
A miniaturization of the micropipette technique used in MGGs, called microfluidic multi-injector (MMI), can generate overlapping gradients using integrated valves for flow injection. This method gives better reproducibility and quantification of gradients than its macroscopic analogue, but is quite slow compared to other μGGs since it requires ~10 minutes to achieve a steady-state gradient.
In general, μGGs are able to generate gradients with much better spatiotemporal resolution than MGGs. The choice of the types of μGGs depends on the constrains of the biological experiment. Flow μGGs are the best candidate for experiments requiring fast response and/or complex gradient shapes. On the other hand, hydrogel- or microchannel-based flow resistive μGGs are best suited for experiments where shear stress, cell drift, and/or the unwanted flushing of secreted cellular factors are of concern.
Figure 2. Microfluidic gradient generator.
(A) Schematic representation of a T-sensor with two inputs. One can see the diffusion between the two laminar streams along the device [24].
(B) µGG composed of an array of 16 multiplexed inlets which allow 64 combinations of gradient generation. This figure shows gradient shape modification in the central chamber depending on the valve state [28].
(C) A microfluidic cell culture array containing 100 cell culture chambers with integrated gradient generators [162].
(D) Zoom on the cell culture array with gradient generation demonstrated using red, blue and yellow dyes [162].
Microfluidics, due to the small time constant and diffusion mixing, can be used to dynamically focus a drug stream on a given part of a cell. This ability allows dynamic observation of cell behavior immediately after drug treatment, or to study how local chemical stimuli propagate in the cell [34, 35].
Figure 3 gives examples of the possibilities offered by laminar flow for cell treatments.
Similar to µGGs, partial treatment of cells using laminar flow methods (PARTCELL) can expose part of an adherent cell situated at the edge of two or more laminar streams of membrane permeable molecules [35, 36], where the width of the laminar mixing region can be between 1μm-10μm.
This method is limited by molecular diffusion since small molecules can diffuse ~100 μm in less than 1 minute, and thus can diffuse throughout the whole cell during PARTCELL treatment. This effect smooths the molecular distribution and creates molecular gradient inside the cell instead of a straight molecular concentration step. Wheeler et al. showed that, by controlling the inlet flow rate, the displacement of the laminar mixing region can be achieved, and thus media switching can occur in less than 100 ms around a cell [37]; which is 10 times faster than standard perfusion chambers. Hersen et al. have used these kinds of chemical oscillating signals to extract kinetic information on the HOG MAP kinase pathway [38].
A second method to expose a part of a cell, called “hydrodynamic focusing”, uses flows from two sides to squeeze a central flow to widths as small as 50 nm [39, 40]. The position of the hydrodynamic focusing can be changed by simply adjusting the inlet flows; and stable focusing position and width are maintained by constant flow rate [41]. This method has low reagent consumption and good spatiotemporal resolution, but requires precise fluid handling systems and is also limited by molecular diffusion on the cell cytoplasm.
Starting from T-sensors, more complex devices have been designed to automate condition variations for cell biology. For example, King et al. designed a method called “flow encoded switching” [42], to simultaneously deliver different temporal profiles of chemical stimuli, such as pulsed train of different widths or frequencies. Sabounchi et al. subjected Hela cells to biochemical reagents in a pulsatile manner, using external solenoid valves [43]. They were able to apply and remove a reagent from the cells in 100 ms.
In addition, the high permeability of PDMS to gas enables control of gas composition in cell cultures. By flowing gas of controlled composition in microchannels adjacent to the cell culture, it is possible to locally control the gas composition of the medium [44, 45]. This kind of device enables fast gas composition switching or gas gradient generation.
The devices presented above are adapted for fast changes in cellular microenvironment. However, in most cases, speed may be of less importance than control flexibility, cell seeding practicality, low shear stress, long-term culture possibilities, and large-scale integration. Substrate patterning / Substrate modification is required for adherent cells. In most of the devices described above, cells adhere on glass surfaces via incubation of the microchannels with ionic polypeptides like poly-lysine or proteins like fibronectin. In cases where cells have to adhere to PDMS surfaces, plasma curing promotes adhesion, but it is not stable over time [46]. To overcome this limitation, it is generally necessary to covalently bind the adherent molecules to the PDMS surface. We will describe techniques to control the chemical nature of the cell substrate.
In standard cultures of adherent cells, cells are randomly seeded on the surface of the culture substrate. This random organization, which is not representative of living tissues, does not interfere with the results of most biological experiments. However, recent advances in cell biology highlighted the need for controlling patterns of cell adhesion to ask questions concerning tissue morphogenesis and cell communication. This leads to the development of three major modes of spatial controls of cell adhesive substrates:
We will next discuss the existing methods to create patterned substrates of various chemical compositions to constrain cells, or to vary other physical properties of the substrate such as patterning of the substrate by molding [51] to study durotaxis behaviors [52].
Figure 3. Spatiotemporal drug control.
(A) Picture from a 2 minute movie showing successive perfusion with Trypan blue dye on a live cell, and subsequent methanol and Trypan after cell death.
(B) Picture showing the ability to change the stream in contact with a cell by changing inlet flow rate. This type of medium switching can be done in 130 ms [37].
(C) Schematic of PARTCELL principle [36]. Using laminar flow properties, partial treatment of a cell can be achieved. Picture shows treatment of a portion of a single cell with Latrunculin A and blue dye.
(D) Picture showing hydrodynamic focusing. Flows B (green) arriving from both sides of flow A (red) focused and maintained flow in a fine stream configuration [40].
The oldest method to pattern biomolecules is photolithography using UV light to expose a mask containing the desired patterns onto photosensitive resists. It is then possible to transfer onto the resist patterns of biomolecules of interest by etching or liftoff. This method has been successfully applied to produce adhesion patterns for cell culture [53]. Nevertheless, it requires clean room facilities which are generally not easily accessible for biologists, and the chemicals used in the process may be harmful for some biomolecules. To avoid this last drawback, a water-soluble sacrificial layer like agar of PVA can be inserted for protection [54].
An alternative to photoresists is the use of UV to directly degrade molecules [46]. By using deep UV to degrade a repellant molecule like PLL-PEG, it is then possible to reactivate the surface for adsorption of the biomolecules of interest afterwards [55]. This method, which works on glass and PDMS substrates, is particularly robust and easy to process. Interestingly, variants of this method can also produce gradients of the surface concentration of certain molecules [56].
Photolithographic methods are probably the best methods in terms of pattern quality and patterns with high resolution (down to 1 μm). However, these methods have to be modified for each new substrate, and are thus not versatile.
Figure 4 gives examples of microcontact printing methods and applications. Microcontact printing (μCP) is a method which enables printing patterns of molecules on a substrate using a microstructured stamp (generally fabricated in PDMS by molding). The stamp is coated with the molecule of interest by dipping it into a solution which can contain a multitude of elements such as thiols, proteins, silanes or nanoparticles. Once the molecule of interest is adsorbed on the stamp, the stamp is temporarily put in contact with the substrate to allow transfer. After printing on a surface, the nonprinted adjacent surface can be made passive with another molecule to prevent cell spreading beyond the printed areas. The PDMS stamp can be used ~100 times over a period of several months without noticeable degradation of the quality of the printing [57]. When using classical PDMS, this technique can achieve resolutions below 500 nm [58]. μCP has been used extensively for substrate patterning of biomolecules for experiments such as axon guidance [59, 60], or cell culture on defined geometry [46]. μCP was initially used to print self-assembled monolayers of alkanthiolate on gold surfaces to perform hydrophobic patterning. This very efficient technique was soon extended to patterning of peptides, proteins, and a wide range of biomolecules on different substrates. Nevertheless, all molecules cannot be stamped using μCP since the “ink” has to be dried on the PDMS stamp to be patterned. To avoid drying, agar stamps can be used [61]; such methods have achieved resolutions down to 50 μm.
When printing of several different molecules is needed, it is possible to perform sequential functionalization of the substrate by using different stamps with different molecules. This method is easy to perform but requires an aligner to control the position of the successive printing. An alternative method consists in loading the stamp with different molecules simultaneously [58], but this method does not generally give precise spatial protein concentration. Crozatier et al. developed a method based on microaspiration to load multiple samples onto a stamp prior to transfer on substrate [62]. In contrast to single μCP methods, multi-molecules µCP requires strong technical know-how.
μCP also enables direct patterning of gradients onto a substrate. Stamps composed of arrays of high-resolution patterns with controlled spacing and density can generate gradients of biomolecules [59]. An original technique involving μCP directly patterned bacteria at cellular resolution on a substrate using structured PDMS stamp with bacteria as the ink [63].
In general, μCP enables high-resolution patterning with a large range of patterns. μCP can pattern planar or non-planar substrates and have been used on substrates such as glass, silicon, and polystyrene. μCP patterning is limited to molecules that are not altered when adsorbed on a substrate, and typically is limited to patterns containing only one or two types of molecules. During an experiment, the adsorbed biomolecules may also degrade or be replaced by other molecules in the medium.
Figure 4. Substrate patterning using microcontact printing (µCP).
(A) Schematic of µCP procedure [58].
(B) Bicolor µCP using successive stamping of molecules [58].
(C) Multicolor µCP using stamp pre-inked with molecular gradient [62].
(D) Influence of adhesive micropattern on cell cytoskeleton. This figure shows vinculin and actin repartition for different fibronectin patterns [50].
An alternative method called “stencil patterning” enables patterning of any component without altering it. This method requires covering the substrate with a membrane (typically PDMS) containing microholes (the stencil). Deposition of molecules is applied to all surfaces, but only the microholes are exposed. Subsequent removal of the membrane stencil produces treated surfaces at the microholes’ positions. This method is often used to locally apply a harmful treatment on a delicate surface [64], or to directly pattern cells onto a homogeneous substrate [65]. Stencil patterning is a convenient method but manipulation and fabrication of micromembranes remains tricky.
Figure 5 gives examples of substrate patterning by flow and active elements. Liquid flow can also achieve patterned substrates. Patterns can be formed by restricting the flow area using microchannels directly on the substrate, or on an intermediary stencil. This method allows for simultaneous deposition of a large number of ligands by circulating different streams in parallel or in a gradient [66]. Liquid flow enables successive treatments on the patterned area, but often the shape and resolution of this method are limited.
To pattern the substrate by liquid flow, one method consists in sticking a microfluidic channel on the substrate and flowing it with a solution containing biomolecules. The microfluidic device can be permanently stuck if the biological experiment is performed in the same device [66], or temporarily stuck by simply putting down the device on the substrate [67], or by using vacuum aspiration [68]. Using this method, Delamarche et al. showed that it was possible to pattern biomolecules on various substrates like glass, gold or polystyrene with submicron resolution [67], and Folch et al. created protein templates of collagen and fibronectin allowing cells to adhere on selective surfaces [69].
By flowing different solutions of biomolecules on different channels in parallel, it is possible to create patterns of multiple compositions or to pattern molecular gradients on a substrate using a μGG [23, 66]. A simpler but less versatile method is to use the depletion effect of the solution while it flows along the channel to generate a gradient [70].
Using PDMS multilayer devices, it is also possible to indirectly pattern cells at desired locations on a substrate [71]. This device uses integrated structured valves to first restrict flows of blocking agents onto surfaces uncovered by the structured valves (Fig. 5). After releasing of the valves, flows containing ECM will adsorb ECM to the unblocked surfaces. This device produces adhesive islands on a microchannel for cell culture.
These described flow patterning techniques generally only pattern continuous shapes if no stencils are used. To overcome this limitation, Chiu et al. proposed a 3D microfluidic device to pattern discontinuous patterns [72]. This device has been used for classical chemical patterning and also for direct patterning of different mammalian cells on the same substrate. The major drawback of this device is the complexity of the fabrication process compared to the other flow patterning techniques.
Figure 5. Substrate patterning using flow or active elements.
(A) Structured valve based microfluidic device for substrate patterning. A1-A2: Schematics showing cross-sectional view of the device and deformation of structured membrane when applying pressure on the top channel. Flow of passive agent recovers only the area unprotected by membrane structure, allowing future adhesion of ECM protein. A3: Picture shows a scanning electron micrograph of the membrane structure. A4: Fluorescence picture of actin stained endothelial cell following ECM matrix protein shape [71].
(B) Direct cell patterning using reversible device with two interconnected channel layers. Up: Schematic representation of the method used to pattern different cell types on the same substrate. Down: Fluorescence picture of two cell types deposited on a tissue culture dish in a concentric pattern using this device [72].
(C) Dynamic cell patterning. Picture showing BCE cells attached to a surface patterned with specific thiols. Application of a cathode voltage pulse allowed release of the cells from the micropattern (time in minutes) [79].
To micropattern a substrate with chemicals and without contacting it, it is possible to use an inkjet printer or a microdroplet dispenser which can deposit a large number of different molecules onto the substrate [73]. This technique is also applicable to directly pattern the cells themselves [74]. Nevertheless, the resolution is limited to several tens of micrometers and the shape of small patterns is generally limited to a disk.
Methods for reversible substrate patterning have been developed to expose cells to a dynamically reversible surface chemistry. In traditional studies, a highly invasive method uses scraping away of cell monolayer to investigate cellular response to the newly exposed ECM [75]. Now, for these kinds of studies, the thermo-responsive polymer PNIPAAm, which changes from hydrophobic to hydrophilic with temperature can be used [76]. Coupled with collagen, this polymer enables control of cell attachment and detachment from the substrate. This polymer can be patterned on substrates using photolithography [77], and can also be coupled with growth factor and insulin to stimulate cells growth. Another promising method involves photosensitive molecules which enable easy switching of cell adhesion directly through the microscope. Nevertheless, sensitivity of photosensitive materials needs to be improved [78]. Electroactive polymers such as self-assembled monolayer (SAM), which can be switched by electrodes, can also be used to control the adhesion state of cells on surfaces [79, 80].
Figure 6 gives examples of topological patterning. In addition to substrate chemistry, substrate topology can also be patterned. The simplest method to integrate structures in PDMS devices consists in sticking a PDMS channel onto another PDMS layer structured with the desired pattern [81], or to directly mold a PDMS replica from a 3D mold containing both the channel and the structure [2]. These kinds of devices can be used to perform force measurement [82], or integrate microwells arrays which enable individual cell culture with easy cell docking by gravity sedimentation [81] or capillary force [83]. The capillary cell deposition technique is faster and more reproducible than sedimentation but exposes cells to air for several seconds which could be a major drawback for cell viability.
It is also possible to structure the glass substrate before bonding to PDMS channels [83]. Traditional approaches to pattern glass used etching procedures but those techniques require strong know-how, facilities and are not easily accessible to small microfluidic labs. An alternative method to structure the glass substrate is micromolding in capillary (MIMIC), which enables the fabrication of 3D structures on the substrate by injecting the material to be molded (sol-gel, salt, polymer beads, colloid, etc…) into a PDMS channel temporarily in contact with the substrate [51]. This method sometimes gives rise to problems of insufficient air evacuation to prevent air bubble in the channel and incomplete molding. Folch et al. solved this problem through injection by aspiration, but this method does not guarantee a complete filling of the channel [65]. Le Berre et al. integrated microaspiration networks around molding channels to completely release all remaining air bubbles [84]. A useful application of MIMIC is substrate patterning using hydrogel, which solidifies inside the microchannel to generate compartments for passive feeding of cells. Lee et al. used microfluidic moldings to fabricate 3D ECM structures of aligned collagen fibers for cultures of endothelial cells [85].
Other major applications of substrate physical patterning are durotaxis, cell guidance and cell growth control. Durotaxis studies use the ability of cells to sense the surface topography and stiffness. Solon et al. showed that fibroblasts tend to adapt their stiffness to the stiffness of substrate [86], and Engler et al. showed that stem cell differentiation is substrate stiffness dependent [87]. Soft-lithography can create PDMS substrate stiffness from 12 kPa to 1000 kPa depending on the ratio of elastomer to the curing agent [88]. In addition, hydrogel can be created with a gradient of crosslinkers to produce a substrate with a gradient of different mechanical stiffnesses, which can be used to study cellular response to substrate rigidity [89]. Substrate physical patterning can also be used to control cell adhesion and growth on a surface with a defined topology inside a microfluidic channel. Depending on the cell type, microstructure integration enables control of cell morphology and motility without the need for chemical stimuli or contact guidance [90, 91]. Such devices with integrated microstructures have been proposed to guide cell movements, cell separation [92], or guidance of neuronal axon growth [93].
Figure 6. Substrate physical patterning.
(A) Top: Schematic of cell docking in microwells using capillary force. Down: Picture of SG3 yeast docking in microwells [83].
(B) Cell guidance using physically modified substrate. Left: Picture of corneal epithelial cells on SiO2 substrate with 70nm wide ridges. Right: Cell on a smooth SiO2 substrate [91].
(C) Example of pattern made by micromoldling technique. From top to bottom: 150nm diameter pillar, fluorescent image of 40µm stripe of quantum dot, 100µm hole in a 20µm thick PDMS layer, optical image of 100µm width and 20µm height wall of agar gel [84].
The behaviors of cells are directly related to their mechanical environment, namely, their confinement or the applied stresses. These environmental constraints have been shown to affect fundamental biological processes such as cell growth, differentiation, ECM metabolism and gene expression. Microtechnology can integrate sensors and actuators with the size of a cell. Depending on the application, the geometry of a cell can be imposed or forces on cells can be applied and measured. The application of these constraints can be used to:
In addition, diseased cells show different mechanical properties than healthy cells. As an example, some cancer cells are more deformable than healthy cells [100]. This deformability has implications in metastasis spreading over tissues and allows automatic sorting of cancer cells by mechanical properties. In this part, we will describe the different methods used to measure and apply strains and stresses on a cell using simple microdevices.
Figure 7 gives examples of microfluidic tools for cell deformation and force measurements. By forcing a cell to enter a small space, one imposes its geometry. For example, Takeuchi et al. used agarose and PDMS microchambers to force E. coli cells to grow in a circular or sinusoidal shape [101], and Terenna et al. used curve PDMS microchannels to constrain growing yeast cells, which are normally straight, to grow bent in order to study the reorganization of its cytoskeleton [94]. Minc et al. achieved similar results by pushing long yeast cells into PDMS microwells, which made the cells buckle and bend [102]. This method has been used to study yeast polarization and to determine elastic modulus of the cell wall and cell turgor pressure [103]. Beyond shape changes, artificial confinement can induce cell behavior which illuminates underlying mechanisms. For example, Faure-Andre et al. injected dendritic cells into microchannels and found that the cells recovered their integrin-independent motile behavior found in living tissues [98]. Chau et al. used a PDMS/matrigel device to study cancer metastasis by following the migration and deformation of cancer cells in microchannels [104]. In these kinds of static cell shape studies, the nature of the walls of the channels may affect cells behavior, and long-term media exchange is also a concern. In another example, channel geometry has also been used to confine microtubules movement on kinesin-coated substrate [105], [106].
To study dynamical changes in cell geometry, active elements such as deformable membranes are required. Kim et al. used a bilayer PDMS device with a membrane to deform cells as a function of the applied pressure [107]. They used this device to perform a mechanical lysis of cells. On the other hand, we used a similar squeezer to study the reconstruction of the cell cortex (Fig 7B, our unpublished data).
Figure 7. Applying and measuring forces on cells using microdevices.
(A) Constraining cells using microchannel: Picture of rod-shaped fission yeast cells in a curved microchannel. Cells have to bend to conform to the channel’s shape [94].
(B) Using microfabricated membrane to deform cell: 3D reconstruction of confocal picture of a cell squeezed by a pressurized PDMS membrane. Increase in membrane pressure leads to cell deformation and cell blebbing. Top: horizontal cross section, Bottom: vertical cross section (Red: membrane, Green: nucleus)
(C) Microfabricated chambers for cell constraint. Left: Schematic of the system showing how yeast cells are placed into the PDMS chambers using the pushing forces of the objective lens. Right: Picture of the cell before (unconstrained) and after entering the microchamber (constrained) [102]. This kind of device can also be used for cell mechanical properties measurement using the deformation of the PDMS chamber due to cell turgor pressure [103].
(D) Measuring force using micropillars array: Scanning EM picture of micropillars array allowing cell force measurement using pillars deformation [82].
In a 2020 Cell paper, co-workers from Sara A. Wickström’s group in Helsinki, used this cell confinement method to study the impact of mechanical stress (stretching and compression) on DNA damage. They showed that cells and tissues protect DNA integrity in a two step mechanism. First, DNA damage is prevented in stem cells through short term nuclear softening and long term monolayer reorientation. Nava, Miroshnikova et al. used a nucleus compression microfluidic model to characterize this first and rapid response (Fig 8). Second, on longer timescales tissues respond to mechanical stress by re-orienting their cytoskeleton to dissipate forces from cell to cell while at the same time restoring baseline heterochromatin [188].
Adherent cells cultured on a flat substrate can only exert forces tangentially to the surface [108]. Existing methods to measure these forces are generally called traction force microscopy, which uses engineered substrates to map the forces exerted on their surfaces. The first experiment of this kind used a substrate composed of a thin film of polymerized fluid silicon layer. The wrinkles of the film due to cell contraction on its surface viewed by phase contrast microscopy gave qualitative direction of forces exerted by the cell [108]. An improved method measured the deformation of a soft substrate by tracking displacement of embedded fluorescent beads [109]. This method enables measurement of the force field under a single cell, but requires heavy computation and image analysis due to propagation of deformation through the substrate. The sensitivity of the measurement depends on the stiffness of the substrate. For example, to measure forces exerted by mammalian motile cells, polyacrylamide hydrogel is generally used due to its low elastic modulus, which is of the order of the pressures applied by the cells (~1 kPa) [110]. The major limitations of these methods are the difficulty in keeping the beads in focus during the experiment, and the need for a reference image containing no cell to calculate beads displacements.
Microfabricated pillars, acting as cantilevers, can be used to measure forces. In this method, an array of micropillars shows local displacements due to the cell grown on top of the pillars. Since the sensitivity of this measurement depends on both the stiffness of the material and the geometry of the micropillars, it is possible to tune the substrate apparent stiffness by modifying the pillars geometry. In addition, since the position of each pillar is known, there is no need for a reference image to calculate forces. And although the spatial resolution remains limited by the pitch of the pillars network, the calculation of forces is straightforward.
Soft-lithography molding of PDMS enables the fabrication of high aspect ratio micropillars which are adapted to the range of the forces exerted by cells [88]. The deflection of such vertical PDMS pillars can measure forces from 50 pN to 100 nN per pillar [82, 111]. In most of these studies, the pillars were in the range of 2-3 μm wide, 6-10 µm high and 4-9 μm spacing apart. The high aspect ratio required for mold fabrication of these pillars generally requires Deep Reactive Ion Etching (DRIE), which remains a technology not easily adapted to most cell biology labs.
In a confined environment, cells can also push or pull on the wall of the substrate. Methods previously mentioned cannot measure side forces. However, it is possible to measure side forces using pillars or thin walls. If the cells move between the pillars or walls, displacements can be converted into forces. This method has low precision and the confinement is not easily controlled. Another method uses PDMS devices with a deformable membrane that can measure and apply forces to cells by controlling the pressure on the membrane sitting on top of the cells while monitoring the membrane deflection. Using this method, Hohne et al. characterized the elastic modulus of objects in the range of 0.1 – 100 kPa. Hohne et al. and Bechinger et al. measured the force produced by a bacteria film and a single fungus appressoria using a PDMS membrane with optical deformation measurement [112, 113]. Discover our review about fungus indentification using microfluidics technology and our protocols on how to perform fungus encapsulation or how to perform fungicide screening. More complicated apparatus have also been used such as atomic force microscope [114], magnetic beads [115], optical tweezers [116], or complex dedicated force machines [117]. Most of these apparatus measure global and local forces of cells. To our knowledge, no method exists to map normal forces applied by a cell in a confined environment with a sub-cellular spatial resolution.
If the precise geometry of the cell is not of concern, cells can be macroscopically subjected to stress. Stress can be compressive, as an example, to study bone cell responses to mechanical stresses [118]; or stress can be stretching, stretching a flexible membrane on which cells are attached [119]. Microfluidics can be used in the application of a shear stress using a flow. Indeed, in a microchannel, cells can easily be submitted to laminar shear or extensional flow [120]. This situation is similar to vascular ducts, and can be used to produce artificial vessels to address questions such as vessel clogging by Sickle blood cells [99]. Flow can be used to probe the adhesive strength of the cells [121], or lyse cells if coupled to nanoscale barbs [122].
Finally, other methods exist to exert forces on cells, such as embedding magnetic micropillars in an elastomeric substrate, and through a magnetic field, exert forces on cells [123], or using optical stretcher to produce cell surface stresses and thus determine cell deformability in microfluidic devices [100].
Temperature is a key parameter for cell biology studies. Biochemical reactions in cells can be temperature-dependent, such as polymerization of microtubules [124]. In addition, genetic model organisms have conditional temperature-sensitive mutations which can be switched on-off by changes in temperature. Due to their small time constant, microfluidics have strong potential for applications involving temperature changes faster than one minute. In this part we will describe the different ways to control temperature during biological experiments involving imaging cells. First we will describe traditional temperature controls and then we will focus on microfluidic temperature controls applied to cell biology.
Microscope incubation box is the most common method used to control the temperature of a biological experiment. The box gives stable and precise (0.1°C resolution) temperature control, but due to its size, requires several tens of minutes to achieve temperature changes. For faster temperature changes, a heated microscope stage, modified Petri-dish with integrated transparent heater, or a Peltier system with a drilled hole in the center to allow microscopic visualization have been developed [125]. These systems allow temperature changes of several degrees within minutes, but they generate strong temperature gradients (2-3°C), particularly when used with immersion objective lenses, which act as a heat sink. These macroscopic methods are generally slow and do not produce homogeneous temperature fields. One current method for fast temperature changes in the seconds range uses microperfusion. However, perfusion introduces shear stresses on adherent cells, and is not adaptable to nonadherent cells.
Some biological applications require fast and precise temperature controls. For example, cytoskeletal dynamics are in the seconds range, and temperature-sensitive mutations can respond within a few minutes. Microfluidic devices reduce the volume needed for temperature control and thus allow faster temperature changes than macroscopic systems.
Figure 9 gives an example of integrated temperature control. A simple microfluidic device using fluid flow can give fast and efficient temperature control. Indeed, laminar flow can generate a strong temperature gradient since heat transfer at the interface remains only diffusive. However, since the temperature diffusive coefficients of most materials are ~100 times higher than the diffusion coefficients of small ions, it is much more difficult for temperature gradients to reach high spatial resolutions reachable by chemical gradients. Nevertheless, Lucheta et al. used the laminar properties of microfluidic flow to generate a temperature step in a microchannel and submit the two sides of a Drosophilia embryo to different temperatures [126]. Pearce et al. used the ability of laminar flow to generate strong temperature gradients to culture neurons and recorded neuronal activities using an integrated electrode array [127]. Groisman et al. used a bilayer microfluidic device to control the temperature of a cell culture chamber by flowing temperature controlled water through the top layer [128]. Our group used a similar device to perform temperature changes of a cell culture in a range of 2°C-40°C in less than 10 seconds (our unpublished data). We used this device to control cytoskeletal dynamic and to activate/deactivate temperature-sensitive gene products during mitosis.
To reach higher spatiotemporal resolution with temperature, it is possible to use integrated electrodes which enable temperature changes of several tens of degrees in less than a few seconds. These kinds of devices have been extensively used for integrated PCR [129]. However, they are not transparent, and thus not usable for experiments requiring microscopic viewing of cells. For cell cultures, transparent heater electrodes can be made of a transparent conductive material such as ITO [130]. However, these kinds of devices cannot cool actively, and require clean room facility, making them limited for general biological usage.
Figure 9. Integrated temperature control.
Temperature step generator.
Left: Schematic of PDMS microfluidic device generating a temperature step around a Drosophila embryo.
Middle: Visualization of the temperature step with thermochromic liquid crystal (green: 21°C – red: 28°C).
Right: Influence of the temperature step on the embryo nuclear density after 150 min [126].
Electric field is widely used in biological research:
Depending on the application, electrodes can be integrated or externalized. The characteristic length of microsystems leads to a favorable scaling of electric fields. It is possible to generate high electric fields with very small voltages.
Compared to optical detection, electronic measurements provide a more convenient form of information recording, and are easier to integrate in a small volume. In addition, most electrical detections do not require specific protein labeling of tagging necessary for optical methods. The main drawback for applications with integrated electrodes is that their fabrication requires clean room facility. In this part, we will discuss the potential of integrating electrochemical and electrical elements into microfluidic devices for cell biology. We will first focus on the use of electric fields for cell handling and electroporation. Then we will discuss about the integration of electrochemical sensors. Other excellent reviews already exist concerning electrical forces for microscale cell handling [131], and electrical analysis of cell in microfluidic systems [132].
Forces applied by electric fields can be used for the handling or the separation of cells. Two electrical phenomena can be used to displace cells: electrophoresis (EP) and dielectrophoresis (DEP). EP interacts with the charges of cells, while DEP interacts with cell polarizability. In both cases, the applied force depends on the cell type, and thus can be used for cell separation.
In contrast to EP for separation of biological molecules, EP to separate cells is seldom miniaturized. Indeed, the main advantage of miniaturizing an EP system is to reduce the dispersion of samples being separated. Cells diffuse slowly enough that macroscopic EP devices are already able to generate the highest separation precision. Nevertheless, microscopic EP can be integrated to transport cells in a chip or to integrate an EP step in a more complex lab-on-chip. As an example, Li et al. used full glass EP chip to transport different cell types like yeast, erythrocyte, and E. coli in a microchannel [133].
DEP is generally considered better than EP for cell separation in microsystems. DEP forces increase when size is reduced, and DEP has few adverse effects on cells [134]. Different cell types have different protein expression patterns which produce different cell polarizability. This phenomenon enables DEP to separate different cells types, even dead cells. DEP can also be used to spatially position cells using stationary field traps which can be dynamically inactivated or moved. This method can potentially create long-term patterns of nonadherant cells with the ability to move them independently. Depending on the type of manipulation, different electrode sizes and configurations can be fabricated, such as an interdigitating electrode array for cell trapping [135], or quadripole electrode array [136].
When a pulse of electric field higher than 1 KV/cm is applied to mammalian cells, the cell membrane is temporarily compromised and generates pores which allow exchanges between the intracellular and extracellular medium. This reversible phenomenon is known as electroporation. Electroporation allows molecules which are normally not able to cross the plasma membrane, such as DNA, drugs, proteins, amino acids or dyes to enter or leave the cell. When far higher voltages are used, the pores do not reseal, resulting in cell inactivation or cell lysis. Macroscopic devices for electroporation currently exist, but microsystems have numerous advantages.
Microsystems enable precise positioning of the electrode in the vicinity of a cell, resulting in high and localized electric fields for electroporation at low voltages. The cell can be immobilized during electroporation [137], or flowed through a channel, or electroporated/lysed during passage through a constricted zone [138]. Cells can be massively electroporated in parallel arrays, with a voltage as low as 1V [139]. Using microchannels, the electrodes can be placed externally, and the electric field can be focused onto the cell by constricting the channel. Cells passing through a constricted channel (200 cells/sec) can be electroporated in such a manner [140].
Another advantage of microfluidic electroporation is the better efficiency of poration due to a better homogeneity and reproducibility of the electric field, and a weaker effect of joule heating induced by the applied voltage [141].
Cell lysis enables the study of cell biochemical content. The main advantages of miniaturization of cell lysis are single-cell operation, and coupling of lysis with other operations such as electrophoresis [142]. Many microfluidic devices miniaturize existing macroscopic techniques of cell lysis. Mechanical lysis using chemical detergent [143] is a simple method, but it remains an invasive method for subsequent chemical analysis. Electrical fields can lyse cells faster, and without the addition of chemicals. Similar to electroporation, microscale lysing devices have the additional advantage of using small voltages [138], and can couple subsequent analysis of single cells. As an example, McClain et al. integrated cell lysis and electrophoresis in the same device to perform fast sequential analysis at a rate of 10 cells/min [144]. El-Ali et al. presented a similar device, and included a stimulus area to study fast cell response processes [145].
If two cells are put into contact, the application of a high intensity electric pulse can lead to membrane rupture and subsequent fusion of cells [146, 147]. Improvement on cell fusion can be achieved using microfluidics. By pairing two kinds of cells on an array using microfluidics, it is possible to perform large scale hetero-fusion with a high yield (>50%) [148]. This tool can be used for production of hybrid cells, cloning, or for studying gene expression.
Current or voltage detection, patch clamp and impedance measurements are often used to determine electrical characteristics of cells. These techniques have been integrated into microfluidic devices and will briefly be reviewed here.
Figure 10 gives examples of microfluidic devices for patch clamping and cell fusion. Patch clamp is a method to measure the activity of a single ion channel on the cell membrane. Traditionally, patch clamp is performed by clamping a glass pipette onto the cell surface using micromanipulators and vibration-free equipment. Several commercial solutions currently exist for patch clamping. However, new techniques coupling patch clamp to microfluidics are being developed. PDMS microchannels can be used for patch clamping. Since microfluidic patch clamping is integrated in the bulk of the microfluidic device there is no need for vibration control and the cell can stay in place for longer time periods.
The main difficulty for microfluidic patch clamp is to reach a seal resistance in the GΩ range. Seo et al. proposed a multi-patch clamp array able to automatically place mammalian cells into the clamp sites by hydrodynamic forces, but obtain a seal resistance of only 140 MΩ [149]. Chen et al. designed microchannels for patch clamp with 1-µm aperture, which could reach 1 GΩ seal resistance on 67% of the cells, similar to results obtained with micropipettes [150].
Figure 10: Patch clamp and cell fusion.
(A) PDMS micro patch clamp. Top: Schematic of the patch clamp device. Middle: Picture of the patch clamp array. The small circle indicates one of the patch sites. Bottom: Zoom on the patch site [149].
(B) Cell pairing and fusion device: schematic and picture show two different cell types pairing before fusion [148].
Amperometric methods can detect small amounts of chemical compounds by measuring electron transfer across an electrode. Amperometry is generally used for the detection of electroactive species released by the cell [151]. Confinement of microelectrodes in thin microchannels can enhance system sensitivity while collecting larger amounts of information. As an example, Hafez et al. used four electrode arrays to map exocytosis with subcellular spatial resolution [152]. Sun et al. used fully transparent chip integrated ITO electrodes to detect exocytosis events of catecholamine [153]. Amatore et al. used integrated platinum electrodes to measure chemical response of macrophage to oxidative stress [154]. Amperometric measurements can be coupled with other microfluidic functionalities. Werdich et al. measured extracellular potentials from single adult murine cardiac myocytes with local application of drug [155]. Son et al. integrated cell lysis of red blood cells upstream of amperometric detection of hemoglobin using an interdigitating electrode array [156].
Impedimetric methods measure the electrical conductivity (inverse of impedence) of the sample, which depends on size, shape, and other characteristics. Integration of microfluidics and electrodes enables impedance measurements of individual cells [157], to detect morphological changes, to follow metabolic activity, to count and sort cells, and to perform density measurements, or even exotic phenomena like germination of spores [158] and detection of bacteria in food samples [159]. Impedance measurement of single cell gives information about cell pathology, and therefore can be used to study pharmaceutical effects on cells. Drawbacks include low specificity and long measurement time, making impedametric less useful for applications such as cell sorting.
Figure 10 gives examples of microfluidic cell culturing systems. Cell culture methods are essential for cell biology experiments. Traditional macroscopic culture methods such as ECM-coated Petri dish poorly mimics in vivo conditions such as growth factor concentration, chemical and mechanical stimuli. In addition, large-scale experiments, which consist in doing a large number of experiments at the same time in different conditions, are often not feasible. As an example, evaporation of the culture medium strongly limits the passage from 384-well plate to 1536-well plate. Microfluidics offers several advantages and open numerous new opportunities for research using cell cultures:
We will review here general applications of microfluidics to cell culture. For more details on specific cell types and microfluidics, see Li et al. [133].
Microfluidic systems produce smaller cell culture chamber volumes (nL) and enables parallelization of chambers. Groisman et al. and Balagadde et al. designed nL chambers for bacteria culture with integrated chemostat and media renewal for long-term colony culture in controlled conditions [128, 160]. Thompson et al. designed a cell culture array with integrated gradient generators to monitor fluorescence gene expression in several independent bacteria colonies in different conditions [161]. Hung et al. developed a cell culture array of 100 microperfusion chambers for Hela cells [162]. Gomez-Sjoberg et al. developed an array of 96 chambers with independent culture conditions that can maintain cell viability for many weeks [163]. Yamada et al. designed a simple microfluidic device which mixes two inlets and redistribute them in stepwise concentration downstream [164].
It is possible to integrate both hydrogel and PDMS into a hybrid technology mimicking the 3D cellular environment [165]. This approach has been applied by Lee et al. to fabricate 3D ECM matrices with aligned collagen fibers for endothelial cell culture [85], and by Tan et al. to create layers upon layers of cells co-cultured with different types of cells [166].
System biology, which focuses on the complete understanding of cellular signaling pathways, could also benefit from microfluidic devices. The ability to perform high-resolution spatiotemporal gradients, partial treatment of cells with different chemicals, coupled with the possibility of highly parallel and automated operations at the single cell level will be of particular interest for this field. The potential of microfluidics for system biology is reviewed by Breslauer et al. [167].
Figure 11: Integrated cell culture system.
(A) Single cell trapping array. Picture of a cell trapping array using hydrodynamic trap and zoom on a single cell trap [168].
(B) Droplet based yeast culture system. Top: This device can generate and store 8000 droplets within a 7mm2 area. Bottom: Picture and schematic illustrating droplet loading (scale 100µm) [177].
Medium renewal is a key parameter for long-term cell experiments. Traditional usage of a sealed glass slide is convenient but not good beyond a few hours. The use of Petri dishes or well plates, when coupled with a perfusion system, extend the experimental time. However, this is only useful for adherent cells, and for cells where factors secreted can be flushed away. Microfluidics may provide promising solutions to medium renewal.
For experiments lasting several days or even weeks, the most common microfluidic devices use active pumping of media. Media can be renewed by diffusion between the feeding channel and the channel containing cells. For nonadherent cells, cells can be trapped using hydrodynamic trap [168], or by geometrically constraining cells [169]. Media can thus flow around the trapped cells. Recently, Charvin et al. proposed a microfluidic device for long-term yeast and bacteria cultures using two layers of PDMS channel separated by a cellulose membrane [170]. Flowing medium through the upper channel allowed nutriments to diffuse through the cellulose membrane to the lower channel containing cell cultures, without disturbing cells with flows. This cell culture system is now commercialized by Warner instruments.
The ability to monitor single cell can decouple the noise coming from population heterogeneity [171]. There are several ways to study individual cells using microfluidic devices. The most straight forward method is to pattern adhesive patches on the substrate as described in the previous section. For nonadherent cells, individual hydrodynamic trap, combined with continuous flow [172] and fast media switching [37], is an elegant way to perform single cell experiments. Using this approach, Di Carlo et al. developed a device to automatically position single Hela cells inside a microfluidic device [168]. Lee et al. designed an array of traps to study cell-cell interaction between two different cell types[173].
Single cell culture devices can also be combined with other microfluidic operations. Zhong et al. developed a device which extracts total mRNA from a single human stem cell to perform gene expression profiling [174]. mRNA studies using single cell microfluidic systems seem to give better results than macroscopic experiments [175]. Microfluidics can be used to perform fluorescent hybridization in situ thanks to the fluidic control for spatial transcriptomics pack.
Another convenient way to perform single cell handling and culture is the use of microfluidic droplet generators. Microfluidics are well fitted for automation and handling of a large number of droplets with controlled volume and composition [176]. Schmitz et al. developed a microfluidic device which was able to generate arrays of thousands of droplets with embedded yeast cells [177], and Chabert et al. used droplet generators to isolate and encapsulate cancer lymphocyte cells from a whole blood sample [178]. Microfluidic droplets have the advantage of complete separation of cells, but cell media cannot be exchanged inside the droplets. This method can also be adapted for encapsulation of single cell, and then measure the small amount of molecules generated by the cell [179].
A large number of technologies and materials such as silicon, PMMA, parylene, and even gelatin [180] currently exist to fabricate microfluidic devices [181]. However, because of the simplicity of the technology and the physical and chemical characteristics of PDMS, soft¬lithography of PDMS has become the most popular technology. From a technological point of view, the ability to fabricate microfluidic devices in a period of hours, without the need for an expensive clean room, is very attractive for cell biologists starting microfluidics.
In addition, PDMS has numerous advantages due to its intrinsic properties. PDMS is biocompatible [182], cheap, transparent (240nm-1100nm), with low autofluorescence [183] and can be molded with a submicron resolution [184]. PDMS replica can be covalently bonded to a glass coverslip using simple plasma treatment to form a sealed microfluidic device. The deformability of PDMS enables leak-proof fluidic connection and easy integration of fluidic valves. PDMS deformability also enables fabrication of mechanical sensors scaled to study cell mechanics [82]. Further, this elastomer is permeable to gas [185], enough to allow O2 exchange on chip for mammalian cell cultures [186].
There are drawbacks of using PDMS in cell biology. PDMS can absorb/adsorb small hydrophobic biomolecules and drugs away from the media [187]. Indeed, the absorption properties of PDMS have been exploited as an extraction matrix to remove traces of organic compounds from solutions. Researchers have noticed adsorption of proteins on the PDMS surface, which can be a major problem for cell biological studies. For experiments on cell signaling and determination of drug dosage response, the use of PDMS can strongly bias the final results. To overcome this adsorption problem, a number of PDMS surface treatments have been developed.
A second major problem of PDMS is its permeability to water vapor, leading to evaporation inside the device. This effect can lead to changes in the medium osmolarity, or complete drying over time. It is generally possible to overcome this problem using hydration channel networks, medium renewal systems, or hygrometry controlled environments. Ideally, PDMS devices should be conditioned several hours before usage to stabilize device hygrometry.
Despite these limitations, PDMS microfluidic devices are widely used in cell biological studies. Therefore, it is necessary to understand the effects of the microscale environment to properly interpret and compare the results obtained in microfluidic devices with the ones using traditional methods. There are significant differences in cell proliferation, glucose consumption, gene expression pattern, and mitosis defect between traditional well plates and PDMS micro cultures [187].
For many years, microfluidic devices were mostly developed first in physics/chemistry labs, then with eventual collaboration with biologists. Those researchers developed important technological advancements (e.g. multilayer molds or 3D microfluidic structures), enhanced the capabilities of microfluidics by integrating various active elements (e.g. valves and electrodes), and created efficient and precise setups for fluid handling (e.g. hydrodynamic focusing, droplet formation). Nevertheless, these technologies and setups seldom take into account the needs of routine biological experiments or technical limitations of most biological labs. During the last few years, we have begun to see the setups of basic microfluidic equipments and facilities directly in cell biology labs, leading to the development and application of microfluidic devices more attuned to biological requirements and limitations. In addition, this strong interdisciplinary interaction creates a feedback loop, where limitations in microfluidics discovered by biologists will spur new research in microfluidics, which in turns will spur new research in biology.
For more reviews about microfluidics, please visit our other reviews here: «Microfluidics reviews». The photos in this article come from the Elveflow® data bank, Wikipedia or elsewhere if specified. Article written by Guilhem Velve-Casquillas, Mael Le Berre, Matthieu Piel, Phong T. Tran and revised by Lauren Durieux.
Email* I hereby agree than Elveflow uses my personal data
Do you want tips on how to best set up your microfluidic experiment? Do you need inspiration or a different angle to take on your specific problem? Well, we probably have an application note just for you, feel free to check them out!
Microfluidic flow cells with embedded optical sensors, have been the most common choice for microenvironment monitoring like pH control.
Microfluidic cell culture is an innovative technology to improve classical lab protocols
Microfluidic systems have an important role to play in the democratization of sperm sorting technologies as well as improving existing techniques...
Live-cell imaging is a non destructive method which focuses on the observation of live cells and is widely used in cellular biology research and biomedical industry. This short review will present some technical aspects and challenges encountered in this field. Nowadays, the most widely used cell culture system is the static culture, where the cells are cultivated inside Petri dishes or multiwall plates. In static culture, the culture medium is supplied in a batch-wised manner.
In this review we will describe some label-free microfluidic techniques to accomplish cell separation and sorting.
Ability to switch drugs in seconds or less in microfluidic chambers allows studying cell response at cells timescale.
PDMS advantages and drawbacks for cell biology are here discussed.
This short note summarizes the methods and techniques employed to perform perfusion for live cell imaging and critical issues encountered.
The first live-cell imaging chambers were designed in the early twentieth century, shortly after mammalian cell culture techniques were developped
Live-cell imaging is a non destructive method which focuses on the observation of live cells and is widely used in cellular biology research and biomedical industry. This short review will present some technical aspects and challenges encounter in this field.
This review presents an overview of the different techniques developed over the last decade to regulate the temperature within microfluidic systems.
Cells growth, gene expression and differentiation are often related to mechanical environment such as confinement or applied stresses.
Chemical gradients play a key role in many biological processes and regulate a number of cellular functions in vivo. Indeed, there are several examples of gradient-dependent phenomena in nature.
Get a quote
Name*
Email*
Message
Newsletter subscription
We will answer within 24 hours
By filling in your info you accept that we use your data.
Collaborations
Need customer support?
Serial Number of your product
Support Type AdviceHardware SupportSoftware Support
Subject*
I hereby agree that Elveflow uses my personal data Newsletter subscription
How can we help you?
Message I hereby agree that Elveflow uses my personal data Newsletter subscription