2 A high transfer time of the fluid along the tube: the time for the replacement of the liquid in a tube or the time to fill the tube while experiment has started.
3 A large consumption of sample if the tube has to be filled with an expensive or rare sample. Dead volume can be calculated as follow: V =L*π*ID²/4 where ID is the internal diameter and L the length of the tube.
4 The internal volume of the tube can change because of its deformation if the tube is moved during manipulation or just when flows are stopped because of the fluid inertia. For low flow rates application, these volume changes can induce unexpected flows or oscillations in the microfluidic chips. These artefacts strongly depend on the configuration of the setup and are difficult to predict, but in general, to limit these effects, the use of large tubing have to be avoided for low flow rate applications.
Compliance of a Microfluidic Tubing
For very low flow rate applications, pressure can deform the tube, leading to a change of its internal volume. This change of volume is often responsible for slow down in pressure equilibration when dead-end tubings are used. The deformation of the tube due to internal pressure can be roughly approximated using the following relations: dV/V=2P/E (for thick walled tubes) or dV/V=P*R/(E*t) (for thin walled tubes), where P is the pressure change value, E is the Young modulus of the tube material, R is the tube radius and t the thickness of the tube (values for common tubes are given below). To limit compliance effects of tubes, small and rigid tubes are preferred and can Syringe Pump Flow Stabilizer Microfluidic Kit But sometimes the high compliance of a soft tubing can also be employed in a sophisticated manner to Syringe Pump Flow Stabilizer Microfluidic Kit often observed when using a syringe pump.
Pressure Drop in a Microfluidic Tubing
For thin capillary, the pressure drop along the tube can become non-negligible and or dominating comparing to the pressure drop in the microfluidic chip. This pressure drop, induced by the fluidic resistance of the tube can be approximated by using Poiseuille law as follow: dP =8Lη.Q/pi.R^4. Where L is the tube length, Q is the flow rate in the tube and η is the viscosity of the fluid. Typical numbers are given in the table below. Fluidic resistances are often used in microfluidic setups to get rid of flow rate instabilities and pressure variations created by droplet generation, or enable low fluidic resistance setups to work with low flow rates.
Advantages & Drawbacks of Standard Microfluidic Tubings
| 5/32 Tubing |
5/32 Tubing |
1/16 Tubing |
1/32 Tubing |
1/16 Capillary |
| Elveflow Tubing Kit |
Luer Lock 7/32 Tubing Kit |
1/16 TYGON Tubing Kit
1/16 PTFE TEFLON Tubing Kit |
1/32 PTFE Tubing Kit |
1/16 Peek Flow Resistance Kit |
| Outer Diameter inch (mm) |
5/32 (4) |
1/16 (1.6) |
1/32 (0.8) |
1/16 (1.6) |
| Inner Diameter inch (mm) |
1/8 (3.2) |
500 µm |
300 µm |
65µm |
| Dead Volume (1) |
450 µL |
20 µL |
7 µL |
0.33 µL |
| Pressure Drop (2)
|
0.12 Pa/(µL/min) |
65 Pa/(µL/min) |
500 Pa/(µL/min) |
17 000Pa/(µL/min) |
| Compliance Factor (3) |
18%/MPa (8.1 µL) |
11%/MPa (220 nL) |
0.4%/MPa (2.8 nL) |
0.05%/MPa (16 pL) |
Maximum
Operating Pressure |
2.5 bar |
> 10 bar* |
> 100 bar* |
> 100 bar* |
| Tubing Type |
Tygon R3603 |
Tygon S54HL or PTFE |
PTFE |
PEEK |
| Advantages |
Compatible with wide range of diameters.
Easy to use.
Negligible fluidic resistance for most applications. |
Easy to handle.
Suitable for most applications. |
Very low compliance.
High pressure resistance. |
Very low compliance.
High pressure resistance. |
| Drawbacks |
Soft tubes (High compliance for low flow rates).
High dead volume.
High volume change if tube moved.
Moderate pressure. |
High fluidic resistance. |
Ferrule is mandatory.
High fluidic resistance. |
Ferrule is mandatory.
High fluidic resistance. |
Typical
Application |
Large flow rate connections (pressure controller- reservoir connexion),
Fluid Transfer. |
Fluid transport.
Syringe Pump Flow Stabilizer Microfluidic Kit
Syringe Pump Flow Stabilizer Microfluidic Kit |
Droplets.
Separative science. |
Separative science.
Microfluidic Flow Resistance. |
(1) for a 10 cm length tube, (2) for a 10 cm length tube,given per µL/min of flow rate for water, (3) volume change for a tube of 10 cm, and a pressure change of 1 bar, * Depends on connexion type.