Results and interpretation:
a) Configuration A: Elveflow OB1 + MUX
– Switch each 4 seconds: Average time= 0.2897
– Switch each 8 seconds: Average time= 0.2933
– Switch each 12 seconds: Average time = 0.2833s
– Switch each 16 seconds: Average time = 0.3007s
– Switch each 24 seconds: Average time = 0.2793s
b) Configuration B: Elveflow OB1
– Switch each 4 seconds: Average time = 0.3337s
– Switch each 8 seconds: Average time = 0.6893s
– Switch each 12 seconds: Average time = 1.0927s
– Switch each 16 seconds: Average time = 1.5757s
– Switch each 24 seconds: Average time = 2.4970s

The results show that the Configuration A (OB1+MUX) can perform a faster medium switch than the Configuration B (OB1). This observation remains true even when the switch frequency time between each of the three tested mediums becomes longer: the Configuration A shows a constant average time whereas the Configuration B shows an increment directly proportional to the switch frequency time (see the figure below).

The reason for this increment is that some flow always comes back from the main channel and passes through the other two channels. The longer this flow carries on flowing the longer it will take to come back through the main channel.
There is an alternative way to perform the medium switch with this configuration which allows to reduce the average switch time: when a little residual laminar flow of the other two mediums is allowed on the edges of the main channel, we avoid any back flow. As a result no extra time will be needed by the medium to come back on every switch.
Conclusion:
As every assay or experiment is different, the best configuration will depend on its objectives and conditions. The following SWOT analysis expects to be helpful to decide which configuration is more suitable in each case.