Single cell encapsulation
At a glance
This application note describes how to encapsulate HeLa cells at the single cell level in picoliter-sized droplets using pressure-driven flow-controlled droplet-based microfluidics. In the described setup, droplets had an average diameter of 51.2 ± 1.2 µm. The procedure covers connecting the OB1 and flow sensors, preparing the oil and cell suspension, generating droplets in a flow focusing chip and collecting the emulsion.
- Observed result
- Experimentally, 90.9% of droplets held 0 cells, 8.7% held 1 cell and 0.4% held 2 or more cells
- Applied pressure
- Oil: 300 mbar; cell suspension: 220 mbar, to obtain approximately 50 µm droplets
- Recommended Poisson parameter
- 0.05 to 0.1, to limit droplets containing two or more cells
- Flow control hardware
- OB1 Mk3+; MFS 2D flow sensor (oil); MFS 3D flow sensor (cell suspension)