This can be used, for instance, in order to improve the manufacturing techniques for producing flow resistance, as it is necessary to find a proper way to measure the flow. However, flow sensors have some limitations when working with extremely low flow values, thus, it can be convenient to run accurate simulations and to compare the data from the sensor and from the simulations in order to reduce the uncertainty of their values.
Experimental measurements
As a validation step or alternative to simulations in the design process of a microfluidic circuit, some experiments can be made to measure flow resistance. In that aim, two magnitudes of the circuit must be measured: pressure and flow rate. To do it, it is necessary to place two pressure sensors, one before and another after the flow resistor, as well as a flow sensor before the resistor.
Once you have those measurements, it is as simple as again using the analogous Ohm’s law to get the flow resistor value between the pressure sensors:

Being P1 the pressure measured by the sensor before the flow resistor (or microfluidic element), P2 the pressure after it, and Qm the measured flow.
Again, for some of those microfluidic circuits, the simulation and experimental data can be compared with the expected values from the theoretical models. In case the similarity is enough for the application, further study of the system, like the variation of some parameters in order to check how the system behaves, can be achieved in a faster way using the electrical-analogous equations.
Conclusions on flow resistance
Flow resistance is clearly one of the most important concepts to which you must pay attention when designing a microfluidic circuit so that you can adjust the operation conditions of your system through the Hagen-Poiseuille equation. The main methods for the estimation of the flow resistance of a fluidic component that have been presented are the following:
- Theoretical models: when possible, they are the simplest solution.
- FEM simulations: for accurate results and complex shapes or just when the previous models are not valid.
- Heuristic methods: to validate the computed models or measure resistance of fluidic components which are too hard to model.
As well as flow resistors, there exists the analogy of flow capacitors, in which some fluid is stored in an elastic or volumetric element where the pressure is stored as potential energy. As it happens with the electrical circuits, these elements can be used to produce delays in the flow response and, together with flow inductive elements, which store kinetic energy, can be used to create fluidic “RLC” circuits [6].
Finally, it can be emphasized that a proper understanding of microfluidic resistance has the potential of allowing a multitude of new applications and design procedures. If a system is created with the necessary constraints for the flow resistors and other analogous elements to properly represent the actual behavior, it would be possible to simulate it in real time, allowing the control of complex microfluidic circuits like the actuators of a soft robot [6]. Additionally, this understanding and new design techniques could lead to a highly automated creation of microfluidic circuits in the same way digital systems are produced today as a combination of small predefined blocks [10].