Cell-cell interactions on chip
At a glance
This application note describes how to fill the two chambers of a Fluidic 688 Interaction Chip with different cell types and apply a controlled flow rate for a dynamic co-culture. In the described setup, pre-stained MCF7 cells in chamber A and HEK293 cells in chamber B showed some cross-chamber infiltration during seeding. The main stages are sequential chamber seeding with Mini Luer plugs, static cell attachment in an incubator, and perfusion with an OB1 MK3+ flow controller and a flow sensor.
- Observed result
- HEK293 cells infiltrated 3mm into chamber A; about 20% of cells at one position in chamber B were MCF7
- Seeding conditions
- 3×10^6 cells.ml^-1 for MCF7 and HEK293; chambers of 37.8 µl each
- Static attachment
- 6 to 24 h in an incubator before perfusion
- Flow hardware
- OB1 MK3+ with at least 1 channel 0/2000 mbar, MFS3 flow sensor, bubble trap