Cross flow membrane for molecular transport
At a glance
This application note describes how to set up a cross flow membrane chip (Fluidic 480 from microfluidic ChipShop) to study molecular transport between two chambers. The main stages are connecting the instruments, purging the tubing of air, running a tracer through one chamber while sampling the other, and analyzing the samples by spectrophotometer. In the described setup, the first tracer arrival at the chamber A outlet was observed after 10 min.
- Observed result
- Plateau after ~18 min indicated concentration equilibration between the two chambers in the described setup
- Sampling sequence
- At 36 µl/min: samples at 5 and 10 min, then one every min; 28 min total
- Analysis
- 400 µl filtered water added per sample; spectrophotometer; tracer-specific calibration needed for concentrations
- Instruments
- OB1 MK3+ (at least 2 channels, 0/2000 mbar), 2 MFS3 flow sensors, 3-way valve, MUX Wire valve controller