Microfluidic resistance calculator
At a glance
This application note describes how to calculate the microfluidic resistance of a device, and the related flow rates and shear stresses, with Elveflow's online calculator. The procedure covers entering fluid properties, computing tubing resistance, defining the channel geometry, then entering a target flow rate, pressure or shear stress. In the demonstration, the calculated shear stress agreed with experimental data in the IBIDI datasheet.
- Observed result
- Demonstration: 0.23 mbar at 1000 µL/min; steady flow hard to achieve without added resistance
- Demonstration channel
- Rectangular IBIDI chip: 5000 um wide, 400 um high, 50 mm long
- Demonstration tubing
- 250 um inner diameter, 8 cm before the chip, entered in the tubing step